Infrared cameras overestimate skin temperature during rewarming from cold exposure
Maley, Matthew J ; Hunt, Andrew P ; Bach, Aaron JE ; Eglin, Clare M ; Costello, Joseph T
Maley, Matthew J
Hunt, Andrew P
Bach, Aaron JE
Eglin, Clare M
Costello, Joseph T
Editors
Other contributors
Affiliation
Epub Date
Issue Date
2020-05-03
Submitted date
Alternative
Abstract
Objective The primary aim of this study was to assess the accuracy of an infrared camera and that of a skin thermistor, both commercially available. The study aimed to assess the agreement over a wide range of skin temperatures following cold exposure. Methods Fifty-two males placed their right hand in a thin plastic bag and immersed it in 8 °C water for 30 min whilst seated in an air temperature of 30 °C. Following hand immersion, participants removed the bag and rested their hand at heart level for 10 min. Index finger skin temperature (Tsk) was measured with a thermistor, affixed to the finger pad, and an infrared camera measured 1 cm distally to the thermistor. Agreement between the infrared camera and thermistor was assessed by mean difference (infrared camera minus thermistor) and 95% limits of agreement analysis, accounting for the repeated measures over time. The clinically significant threshold for Tsk differences was set at ±0.5 °C and limits of agreement ±1 °C. Results As an average across all time points, the infrared camera recorded Tsk 1.80 (SD 1.16) °C warmer than the thermistor, with 95% limits of agreement ranging from −0.46 °C to 4.07 °C. Conclusion Collectively, the results show the infrared camera overestimated Tsk at every time point following local cooling. Further, measurement of finger Tsk from the infrared camera consistently fell outside the acceptable level of agreement (i.e. mean difference exceeding ±0.5 °C). Considering these results, infrared cameras may overestimate peripheral Tsk following cold exposure and clinicians and practitioners should, therefore, adjust their risk/withdrawal criteria accordingly.
Citation
Maley, M.J., Hunt, A.P., Bach, A.J.E., Eglin, C.M. and Costello, J.T. (2020) Infrared cameras overestimate skin temperature during rewarming from cold exposure, Journal of Thermal Biology, https://doi.org/10.1016/j.jtherbio.2020.102614
Publisher
Journal
Research Unit
PubMed ID
PubMed Central ID
Embedded videos
Type
Journal article
Language
en
Description
This is an accepted manuscript of an article published by Elsevier in Journal of Thermal Biology on 03/05/2020, available online: https://doi.org/10.1016/j.jtherbio.2020.102614
The accepted version of the publication may differ from the final published version.
Series/Report no.
ISSN
0306-4565