Abstract
In this paper we report a unique approach to measuring oxygen saturation levels by utilising the wavelength of the haemoglobin instead of the conventional absorption difference. Two experiments are set up to measure the wavelength of the haemoglobin bound to oxygen at different oxygen saturation levels with the help of a spectrometer. We report a unique low cost and robust wavelength monitoring SpO2 sensor that measures the SpO2 by using the colour of the blood and not the absorption difference of oxyhaemoglobin and deoxyhaemoglobin. With use of a spectrometer, we show that the wavelength of the oxygen-bound haemoglobin has a relation to the oxygen saturation level. The proposed device is designed and experimentally implemented with a colour sensor to measure the SpO2 level of the blood.Citation
Cohen, Z. J. V., Haxha, S., and Aggoun, A. (2016) 'Pulse oximetry optical sensor using oxygen-bound haemoglobin', Optics Express, 24 (9)Publisher
The Optical Society (OSA)Journal
Optics ExpressAdditional Links
https://www.osapublishing.org/abstract.cfm?URI=oe-24-9-10115Type
Journal articleLanguage
enISSN
1094-4087ae974a485f413a2113503eed53cd6c53
10.1364/OE.24.010115
Scopus Count
Collections
The following licence applies to the copyright and re-use of this item:
- Creative Commons
Except where otherwise noted, this item's license is described as https://creativecommons.org/CC BY-NC-ND 4.0